FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR(R) PLUS DEFIBRILLATOR

MDR report key: 3922844 · Received July 9, 2014

Report

Report Number
3015876-2014-00786
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
June 13, 2014
Report Date
June 13, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K033275
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TWO DAMAGED INTERNAL HLC BATTERIES, DESIGNATORS BT2 AND BT3. THE DAMAGED HLC BATTERIES COULD NO LONGER TAKE A CHARGE. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DEVICE WAS DISPLAYING ALL THREE ICONS (ATTENTION, SERVICE WRENCH AND CHARGE-PAK) AND WOULD NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
400008 LIFEPAK CR(R) PLUS DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC CRPLUS

Patients

Seq Age Sex Outcome Treatment
1