FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK CR(R) PLUS DEFIBRILLATOR
MDR report key: 3922844
·
Received July 9, 2014
Report
- Report Number
- 3015876-2014-00786
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- June 13, 2014
- Report Date
- June 13, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K033275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TWO DAMAGED INTERNAL HLC BATTERIES, DESIGNATORS BT2 AND BT3. THE DAMAGED HLC BATTERIES COULD NO LONGER TAKE A CHARGE. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEIR DEVICE WAS DISPLAYING ALL THREE ICONS (ATTENTION, SERVICE WRENCH AND CHARGE-PAK) AND WOULD NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400008 | LIFEPAK CR(R) PLUS DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | CRPLUS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |