FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3922826
·
Received February 5, 2014
Report
- Report Number
- 1314492-2014-06299
- Event Type
- Malfunction
- Date Received
- February 5, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 6, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYMPTOM WHICH WAS NOT REPRODUCED. THE DEVICE WAS TESTED AND NO MALFUNCTION COULD BE IDENTIFIED. SYSTEM ERROR 322 WAS CONFIRMED THROUGH THE EVENT HISTORY LOG. THE EVALUATION WAS UNABLE TO DETERMINE THE CAUSE. THE UPPER AND LOWER AUXILIARY ARE KNOWN CONTRIBUTORS TO THIS SYMPTOM AND HAVE BEEN REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322, PER THE HISTORY LOG. ANY PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 74723 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |