INRATIO PT/INR TEST STRIPS
Report
- Report Number
- 2027969-2014-00523
- Event Type
- Malfunction
- Date Received
- June 2, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 19, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
INVESTIGATION/CONCLUSION: IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. ROOT CAUSE COULD NOT BE DETERMINED FROM THE INFO PROVIDED BY THE CUSTOMER. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, A REVIEW OF IN-HOUSE TESTING DATA WAS PERFORMED. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCT PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED. THE LOT MET SPECIFICATIONS AND NO NON-CONFORMANCES WERE DOCUMENTED. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. BASED ON THE INFO AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT LOW INRATIO INR RESULT IN COMPARISON TO THE LABORATORY INR RESULT. RESULTS ARE AS FOLLOWS: DATE: (B)(6) 2014, INRATIO INR: 3.0, LABORATORY INR: 6.5. THERAPEUTIC RANGE: 2.5 - 3.0. THE TIME BETWEEN TESTING WAS APPROXIMATELY THIRTY (30) MINUTES. THERE WAS NO REPORTED ADVERSE PT SEQUELA. THERE WAS NO ADDITIONAL INFO PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320897 | INRATIO PT/INR TEST STRIPS | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 340746 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INRATIO MONITOR SN UNK| WARFARIN |