FDA Adverse Event Malfunction Summary report: N

POLYAXIAL DRIVER, REFORM

MDR report key: 3922458 · Received June 3, 2014

Report

Report Number
3005739886-2014-00029
Event Type
Malfunction
Date Received
June 3, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
PRECISION SPINE, INC.
Product Code
HXX
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE PROBLEM ALREADY KNOWN, NO EVALUATION NECESSARY. METHOD: THE ACTUAL DEVICE INVOLVED IN THIS INCIDENT WAS NOT EVALUATED AS THE FAILURE WAS ALREADY A KNOWN FAILURE AND CORRECTIVE ACTIONS HAVE BEEN INITIATED. CONCLUSION: PREVIOUS EVALUATION IDENTIFIED THAT ALTHOUGH THE EXACT ROOT CAUSE COULD NOT BE CONFIRMED, IT IS POSSIBLE THAT THE DRIVER MAY NOT HAVE BEEN COMPLETELY SECURED INTO THE PEDICLE SCREW. THIS WOULD MINIMIZE THE AMOUNT OF LOAD-SHARING INHERENT IN THE DESIGN. IN EFFORT TO PREVENT RECURRENCE OF ISSUES OF THIS NATURE A DRIVER TIP CHANGE WAS INITIATED TO RELIEVE INHERENT STRESS RISERS IN THE DRIVER TIP. IN ADDITION, THE PLASTIC OUTER SLEEVE OF THE ASSEMBLY HAS BEEN MODIFIED TO MITIGATE THE OCCURRENCE OF THE USER UNINTENTIONALLY UNTHREADING THE OUTER SECURE SHAFT FROM THE TULIP. THIS CHANGE CONSISTS OF OVER SIZING THE PLASTIC SLEEVE, SO WHEN TIGHTLY GRIPPED BY THE USER THE LOAD WILL NOT BE SO INCLINED AS TO BE TRANSFERRED TO THE OUTER SECURE SHAFT, THUS UNTHREADING THE SECURE SHAFT FROM THE TULIP AND LOOSENING THE DRIVER AND PEDICLE SCREW ASSEMBLY DURING INSERTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE PERFORMED ON (B)(6) 2104, THE TIP OF THE REFORM POLYAXIAL DRIVER BROKE. THE BROKEN PIECE WAS RETRIEVED FROM THE HEAD OF THE SCREW AND THE PROCEDURE WAS COMPLETE WITH A SECOND DRIVER THAT WAS READILY AVAILABLE. NO PT INJURY OR DELAY TO THE PROCEDURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323294 POLYAXIAL DRIVER, REFORM SCREWDRIVER HXX PRECISION SPINE, INC. 3055MMR

Patients

Seq Age Sex Outcome Treatment
1