FDA Adverse Event
Malfunction
Summary report: N
2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLUS
MDR report key: 3922251
·
Received May 23, 2014
Report
- Report Number
- 2937457-2014-00886
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- May 1, 2014
- Report Date
- May 1, 2014
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- KDI
- PMA / PMN Number
- K994267
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.
Description of Event or Problem · 1
A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THE FACILITY TECHNICIAN REPLACED THE AIR SEPARATOR AND THE MACHINE WAS RETURNED TO SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307144 | 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLUS | KDI | FRESENIUS MEDICAL CARE NORTH AMERICA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK SALINE (DISCARDED-NOT USED) |