FDA Adverse Event Malfunction Summary report: N

TANDEM T: SLIM INSULIN DELIVERY SYSTEM

MDR report key: 3922138 · Received May 30, 2014

Report

Report Number
3007981285-2014-00873
Event Type
Malfunction
Date Received
May 30, 2014
Date of Event
May 5, 2014
Report Date
May 1, 2014
Manufacturer
TANDEM DIABETES
Product Code
LZG
PMA / PMN Number
K111210
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING FURTHER TROUBLESHOOTING IT WAS IDENTIFIED PATIENT IS USING APIDRA. CUSTOMER TECHNICAL SUPPORT HAS NOTIFIED REPORTER THAT APIDRA INSULIN IS CONTRAINDICATED FOR USE WITH T: SLIM.

Description of Event or Problem · 1

RECEIVED INFORMATION STATING PATIENT WAS EXPERIENCING HIGH BG'S (300 MG/DL).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319822 TANDEM T: SLIM INSULIN DELIVERY SYSTEM INSULIN PUMP LZG TANDEM DIABETES 004628

Patients

Seq Age Sex Outcome Treatment
1 32 YR