SMALL BATTERY DRIVE
Report
- Report Number
- 8030965-2014-10376
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Report Date
- June 24, 2014
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO A MOTOR ASSEMBLY OR ELECTRONIC CONTROL UNIT (ECU) FAILURE DUE TO USAGE WEAR OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE SMALL BATTERY DRIVE DEVICE WOULD NOT RUN. IT WAS FURTHER CLARIFIED THAT THERE WAS NO SOUND OR MOVEMENT ON THE DEVICE. A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO REPORTED PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY