FDA Adverse Event Malfunction Summary report: N

POSEY SITTER ELITE

MDR report key: 3922114 · Received May 30, 2014

Report

Report Number
2020362-2014-00210
Event Type
Malfunction
Date Received
May 30, 2014
Report Date
May 7, 2014
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORTED ISSUE. THE UNIT DOES NOT SEND A SIGNAL TO ACTIVATE THE RED LED. WHEN THE NURSE CALL RECEPTACLE IS WIGGLED, THE NURSE CALL RED LED TURNS ON AND OFF. THERE IS PHYSICAL DAMAGE TO THE NURSE CALL RECEPTACLE. ADDITIONALLY, UNIT HAS BEEN EXPOSED TO MOISTURE AND THE BATTERY DOOR IS MISSING. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED THE ALARM IS SOUNDING, HOWEVER, IT DOES NOT SEND A SIGNAL TO THE NURSE CALL STATION. THE DATE OF THE EVENT IS UNKNOWN. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
320094 POSEY SITTER ELITE KMI J. T. POSEY CO. 8345 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK NURSE CALL CABLE, MODEL 8282, LOT # UNK