FDA Adverse Event
Malfunction
Summary report: N
POSEY SITTER ELITE
MDR report key: 3922114
·
Received May 30, 2014
Report
- Report Number
- 2020362-2014-00210
- Event Type
- Malfunction
- Date Received
- May 30, 2014
- Report Date
- May 7, 2014
- Manufacturer
- J. T. POSEY CO.
- Product Code
- KMI
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
RESULTS: EVALUATION OF THE RETURNED PRODUCT CONFIRMED THE REPORTED ISSUE. THE UNIT DOES NOT SEND A SIGNAL TO ACTIVATE THE RED LED. WHEN THE NURSE CALL RECEPTACLE IS WIGGLED, THE NURSE CALL RED LED TURNS ON AND OFF. THERE IS PHYSICAL DAMAGE TO THE NURSE CALL RECEPTACLE. ADDITIONALLY, UNIT HAS BEEN EXPOSED TO MOISTURE AND THE BATTERY DOOR IS MISSING. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED THE ALARM IS SOUNDING, HOWEVER, IT DOES NOT SEND A SIGNAL TO THE NURSE CALL STATION. THE DATE OF THE EVENT IS UNKNOWN. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 320094 | POSEY SITTER ELITE | KMI | J. T. POSEY CO. | 8345 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | NURSE CALL CABLE, MODEL 8282, LOT # UNK |