FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3922098 · Received July 9, 2014

Report

Report Number
2955842-2014-04184
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
June 12, 2014
Report Date
June 12, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION OBSERVED THAT THE PITCH CABLE WAS FOUND TO BE FRAYED AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE FRAYED STRANDS STUCK OUT AT THE WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S MAIN TUBE EXHIBITED SCRATCH MARKS/ABRASIONS. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S PITCH CABLE AND/OR MAIN TUBE FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE, THE CUSTOMER NOTED THAT THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WAS NOT WORKING PROPERLY, HAD TO BE REPLACED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401138 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10131121 709

Patients

Seq Age Sex Outcome Treatment
1