MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04181
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 10, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE DA VINCI SYSTEM AND IT PASSED RECOGNITION AND ENGAGEMENT TESTING. AN ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS THAT THE INSTRUMENT'S MAIN TUBE EXHIBITED SCRATCH MARKS/ABRASIONS. THE DISTAL END OF MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S MAIN TUBE FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE, THE CUSTOMER NOTED THAT THE ROBOT NEVER REGISTERED THE MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401137 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10121024 201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |