FDA Adverse Event Malfunction Summary report: N

HAKIM PROGRAMMING VALVE

MDR report key: 3921952 · Received July 9, 2014

Report

Report Number
1226348-2014-11804
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
June 19, 2014
Manufacturer
CODMAN & SHURTLEFF
Product Code
JXG
PMA / PMN Number
PK020667
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

IT HAS BEEN COMMUNICATED THAT THE DEVICE AND/OR LOT INFORMATION IS NOT AVAILABLE FOR EVALUATION. WITHOUT THE DEVICE AND/OR LOT INFORMATION IT IS NOT POSSIBLE FOR CODMAN TO CONDUCT A PROPER INVESTIGATION. IF AT SOME POINT THE DEVICE AND/OR LOT INFORMATION DOES BECOME AVAILABLE, THIS COMPLAINT WILL BE RE-OPENED, EVALUATED AND A FOLLOW UP REPORT WILL BE FILED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CONSIDERED CLOSED.

Additional Manufacturer Narrative · 1

UPON COMPLETION OF INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED. COMPLAINT SAMPLE NOT RETURNED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER CONFIRMED THAT THE 2ND VALVE THAT WAS IMPLANTED WAS NOT REPLACED. CUSTOMER CONFIRMED THAT THE FIRST VALVE WAS IMPLANTED AND EXPLANTED INTRAOPERATIVLY. CUSTOMER DID NOT CONFIRM IF THE LUMBAR CATHETER WAS A CODMAN DEVICE, HOWEVER, IT IS LIKELY THAT THIS DEVICE IS NOT A CODMAN DEVICE THE CUSTOMER CONNECTED A LUMBAR CATHETER TO THE VALVE AND IS CONSIDERED OFF LABEL.

Description of Event or Problem · 1

L-P SHUNT FOR THE PATIENT WITH BRAIN HEMORRHAGE WAS CONDUCTED ON (B)(6) 2014. WHEN THE SURGEON CONNECTED THE LUMBAR CATHETER TO THE VALVE AND TRIED PUMPING, THE SURFACE OF DOME GOT BACK TO THE ORIGINAL POSITION SO SLOWLY AND THE OCCLUSION AT THE LUMBAR SIDE WAS SUSPECTED. WHEN PULLING BACK THE LUMBAR CATHETER ABOUT 1 CM, THE FLUID FLOWED FROM THE DISTAL END OF THE LUMBAR CATHETER WITHOUT PROBLEM. THE SURGEON RECONNECTED IT TO THE VALVE AND PUMPED AGAIN, BUT IT WAS NOTED SAME PHENOMENON WITH THE DOME. HE SUSPECTED THE DEFECT WITH THE INTEGRATED PART OF CATHETER TO THE VALVE. HE REPLACED THE DEVICE WITH THE NEW PRODUCT ((B)(4)) AND COMPLETE THE CASE THOUGH THE SAME FAILURE WAS NOTED AGAIN WITH THE REPLACED VALVE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
400817 HAKIM PROGRAMMING VALVE SHUNT, CENTRAL NERVOUS SYSTEM & COMPS JXG CODMAN & SHURTLEFF

Patients

Seq Age Sex Outcome Treatment
1