FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3921873
·
Received May 29, 2014
Report
- Report Number
- 1720753-2014-04556
- Event Type
- Malfunction
- Date Received
- May 29, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 29, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GIB (GENERATOR INTERFACE PCB) AND FFB (FLUORO FUNCTIONS PCB) WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM LOCKED UP WITH A COMMUNICATION FAILURE ERROR. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 316254 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |