INSIGNIA
Report
- Report Number
- 2124215-2014-10548
- Event Type
- Injury
- Date Received
- July 9, 2014
- Date of Event
- April 23, 2014
- Report Date
- May 14, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
ADDITIONAL INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED. THE DEVICE AND COMPETITOR RIGHT VENTRICULAR (RV) LEAD WERE EXPLANTED AND WAS REPLACED WITH NO ISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE WAS RETURNED FOR RELIABILITY ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER EXHIBITED NOISE BEING OVERSENSED. THERE WERE VENTRICULAR TACHYCARDIA STORED IN ELECTROGRAM (EGM). THE PATIENT HAD ATRIOVENTRICULAR NODE ABLATION. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THAT THE CAUSE OF DOUBLE COUNTING WAS DUE TO NOISE. ADDITIONAL INFORMATION INDICATED THAT THE PATIENT WAS ADMITTED FOR LEAD REVISION AND DEVICE REPLACEMENT. HOWEVER, IT WAS DETERMINED THAT THE VEIN WAS OCCLUDED SO AT THAT POINT FURTHER SURGICAL INTERVENTION WAS ABANDONED. UPON FURTHER ISOMETRIC EVALUATION THE ASYSTOLE WAS REPRODUCED GREATER THAN 2 SECONDS AND IT WAS IDENTIFIED THAT THERE WAS EITHER AN INSULATION BREACH OR CONDUCTOR ISSUE ON COMPETITOR'S RV LEAD. PROGRAMMING CHANGES WERE DONE AND IMPEDANCE RETURNED TO 550 OHMS. THE PATIENT WILL THEN BE MONITORED AND WAS SCHEDULED FOR FULL SYSTEM EXTRACTION. THE SYSTEM REMAINS IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400155 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | 1297 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| L| R | 1297| 5076 |