FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3921817
·
Received May 29, 2014
Report
- Report Number
- 1720753-2014-04538
- Event Type
- Malfunction
- Date Received
- May 29, 2014
- Date of Event
- May 6, 2014
- Report Date
- May 29, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE MOTHERBOARD CONNECTION WAS EVALUATED AND RESEATED. THE HARD DRIVE WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE WAS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 316247 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |