FINELINE II
Report
- Report Number
- 2124215-2014-13638
- Event Type
- Injury
- Date Received
- July 9, 2014
- Date of Event
- April 25, 2014
- Report Date
- April 25, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- DTB
- PMA / PMN Number
- P960004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE LEAD WAS SURGICALLY ABANDONED AND REPLACED. AS THE LEAD WILL NOT BE RETURNED FOR ANALYSIS, THE CLINICAL OBSERVATIONS CANNOT BE CONFIRMED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT OF RANGE PACING IMPEDANCE MEASUREMENTS. AN INVASIVE PROCEDURE WAS SUBSEQUENTLY PERFORMED AND THE LEAD WAS SURGICALLY ABANDONED AND REPLACED. ADDITIONAL INFORMATION OBTAINED FROM A FIELD REPRESENTATIVE INDICATED THAT AN X-RAY SHOWED NO VISIBLE LEAD FRACTURE OR DISLODGEMENT. ADDITIONALLY, THE PACE IMPEDANCE MEASUREMENTS WERE OUT OF RANGE IN BOTH BIPOLAR AND UNIPOLAR CONFIGURATIONS AND THE PACING THRESHOLD MEASURED MORE THAN 6.5 VOLTS. FURTHERMORE, THE OUT OF RANGE MEASUREMENT WAS OBSERVED AS WELL THROUGH A PACING SYSTEM ANALYZER (PSA). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 399976 | FINELINE II | IMPLANTABLE LEAD | DTB | CPI - DEL CARIBE | 4470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization| L| R | 4470| 4457| 4469| S602 |