FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3921698 · Received May 29, 2014

Report

Report Number
1720753-2014-04540
Event Type
Malfunction
Date Received
May 29, 2014
Date of Event
May 7, 2014
Report Date
May 29, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. ONSITE INVESTIGATION REVEALED THAT NO CAUSE COULD BE DETERMINED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED X-RAY SECURITY SWITCH ERROR MESSAGES. A REBOOT WAS UNSUCCESSFUL AND ANOTHER SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
316317 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1