ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19356
- Event Type
- Injury
- Date Received
- July 9, 2014
- Report Date
- June 30, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/06/2015 WITH THE FOLLOWING FINDINGS: UNABLE TO DUPLICATE THE COMPLAINT, THE PUMP INFORMATION FOR THE COMPLAINT DATE HAS BEEN OVERWRITTEN. THE BLACK BOX STARTS ON (B)(6) 2015. DUE TO CONTINUOUS USE OF THE PUMP THE BLACK BOX DATA AND HISTORIES FOR THE EVENT HAVE BEEN OVERWRITTEN. NO ALARMS RELATED TO THE COMPLAINT IN CURRENT ALARM HISTORY. AVAILABLE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED BASAL RATES. A DELIVERY ACCURACY TEST WAS PERFORMED; PUMP PASSED AND WAS FOUND TO BE DELIVERING WITHIN THE REQUIRED RANGE. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT IS CRACKED BELOW THE BUMPER PAD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT HAD A BLOOD GLUCOSE (BG) BETWEEN 250 AND 400 MG/DL, WITH SYMPTOMS SUCH AS EXTREME DROWSINESS/DIFFICULTY WAKING UP/CONFUSION. THE REPORTER STATES THAT PUBERTY IS THE CAUSE OF THE BG EXCURSION AND THERE IS NO ALLEGATION OF PUMP MALFUNCTION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT EXPERIENCED HYPERGLYCEMIA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400528 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR | Life Threatening |