FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3921581 · Received February 6, 2014

Report

Report Number
1314492-2014-06567
Event Type
Malfunction
Date Received
February 6, 2014
Date of Event
January 1, 2014
Report Date
January 10, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER REC'D AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYMPTOM WHICH WAS NOT REPRODUCED. THE DEVICE WAS TESTED AND NO MALFUNCTION COULD BE IDENTIFIED. SYSTEM ERROR 322 WAS CONFIRMED THROUGH THE EVENT HISTORY LOG. THE EVALUATION WAS UNABLE TO DETERMINE HISTORY LOG. THE EVALUATION WAS UNABLE TO DETERMINE THE CAUSE. THE UPPER AND LOWER AUXILIARY ARE KNOWN CONTRIBUTORS TO THIS SYMPTOM AND HAVE BEEN REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
76565 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1