FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3921479 · Received July 9, 2014

Report

Report Number
2531779-2014-19391
Event Type
Malfunction
Date Received
July 9, 2014
Report Date
July 1, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 08/22/2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED; THE REPORTED EVENT WAS DUPLICATED. UNRELATED TO THE REPORTED EVENT, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED BELOW THE GRIP PAD. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. REPORTEDLY, THE DISPLAY WAS DIM FOR A COUPLE OF MONTHS BUT STILL CAN BE READ/MANEUVER SAFELY. IT WAS REPORTED THAT THERE WAS NO EVIDENCE OF MOISTURE OR CORROSION IN THE PUMP AND CONTRAST RESET DID NOT RESOLVE THE ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401110 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 32 YR