FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3921456
·
Received February 6, 2014
Report
- Report Number
- 1314492-2014-06560
- Event Type
- Malfunction
- Date Received
- February 6, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 10, 2014
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION COULD NOT REPRODUCE THE REPORTED SYMPTOM. REVIEW OF THE EVENT HISTORY LOG CONFIRMED THE REPORTED SYMPTOM. THE EVALUATION FOUND THAT THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE FAILING, WHICH ARE KNOWN CONTRIBUTORS TO THE REPORTED SYMPTOM. THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FORM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP ALARMED SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 78610 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |