FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3921450 · Received February 6, 2014

Report

Report Number
1314492-2014-06553
Event Type
Malfunction
Date Received
February 6, 2014
Date of Event
January 1, 2014
Report Date
January 10, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS PERFORMED. THE EVALUATION CONFIRMED AND REPRODUCED THE SYSTEM ERROR 105 CAUSED BY A FAILING MOTOR. THE EVALUATION ALSO FOUND 5 OF 6 MOTOR SCREWS BROKEN. THE MOTOR ASSEMBLY AND MOTOR SCREWS WERE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A DEVICE WAS ALARMING SYSTEM ERROR 105. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78609 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1