FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3921446 · Received February 6, 2014

Report

Report Number
1314492-2014-06549
Event Type
Malfunction
Date Received
February 6, 2014
Date of Event
January 1, 2014
Report Date
January 10, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVAL WAS PERFORMED. EVAL CONFIRMED THE REPORTED SYMPTOM OF SYSTEM ERROR 322 THROUGH THE HISTORY LOG, BUT COULD NOT REPRODUCE IT. REVIEW OF THE KNOWN CONTRIBUTORS LED TO THE DETERMINATION THAT THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE THE FAILING COMPONENTS. THE UPPER AND LOWER AUXILIARY ASSEMBLIES WERE REPLACED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED FOR SYSTEM ERROR 322. IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78536 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1