- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. ON SITE INVESTIGATION REVEALED THAT NO CAUSE COULD BE DETERMINED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED 'PRE-CHARGE VOLTAGE ERRORS' UPON SYSTEM START UP. THIS ERROR IS LIKELY TO PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.