SPECTRUM INFUSION PUMP

MDR report key: 3921401 · Received February 6, 2014

Report

Report Number
1314492-2014-06538
Event Type
Malfunction
Date Received
February 6, 2014
Date of Event
December 1, 2013
Report Date
January 10, 2014
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Removal / Correction Number
1314492-080712-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVAL WAS PERFORMED. THE EVAL CONFIRMED AND REPRODUCED THE DISCOVERED SYMPTOM OF A STUCK KEYPAD KEY. EVAL FOUND ALL KEYS INOPERABLE EXCEPT THE ON/OFF KEY DUE TO A STUCK 2ND SOFT KEY, AND WAS CAUSED BY A FAILED KEYPAD. THE KEYPAD WAS REPLACED. BAXTER HAS INITIATED A CAPA TP FURTHER INVESTIGATE THIS ISSUE.

Description of Event or Problem · 1

DURING BAXTER'S DEVICE EVAL, IT WAS DISCOVERED THAT A SPECTRUM PUMP HAD A STUCK KEY. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78600 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1