FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 3921347 · Received May 21, 2014

Report

Report Number
1720753-2014-04476
Event Type
Malfunction
Date Received
May 21, 2014
Date of Event
May 8, 2014
Report Date
May 23, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE MAINFRAME GENERATOR BATTERIES WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED 'PRE-CHARGE ERRORS' UPON SYSTEM START UP. THIS ERROR IS LIKELY TO PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303495 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1