FDA Adverse Event Malfunction Summary report: N

TRILOGY O2

MDR report key: 3921318 · Received May 21, 2014

Report

Report Number
2518422-2014-00838
Event Type
Malfunction
Date Received
May 21, 2014
Date of Event
April 30, 2014
Report Date
April 30, 2014
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION ALLEGING "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
304112 TRILOGY O2 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1040004

Patients

Seq Age Sex Outcome Treatment
1