FDA Adverse Event
Malfunction
Summary report: N
TRILOGY O2
MDR report key: 3921318
·
Received May 21, 2014
Report
- Report Number
- 2518422-2014-00838
- Event Type
- Malfunction
- Date Received
- May 21, 2014
- Date of Event
- April 30, 2014
- Report Date
- April 30, 2014
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K093905
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION ALLEGING "SERVICE REQUIRED" ALARM CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 304112 | TRILOGY O2 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | 1040004 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |