FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3921203
·
Received February 6, 2014
Report
- Report Number
- 1314492-2014-06643
- Event Type
- Malfunction
- Date Received
- February 6, 2014
- Date of Event
- December 1, 2013
- Report Date
- January 10, 2014
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF (B)(4). THE IDENTIFIED ¿DOWNSTREAM OCCLUSION¿ ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP WILL BE SENT. THE DOWNSTREAM TUBING GUIDE, FORCE SENSOR GASKET, AND FORCE SENSOR PUSHER WERE REPLACED.
Description of Event or Problem · 1
DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 136 OCCURRENCES OF "DOWNSTREAM OCCLUSION" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION IS UNKNOWN SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76564 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |