NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2014-00607
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Report Date
- June 12, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF RESUSCITATOR WAS RETURNED TO OUR SERVICE CENTRE IN (B)(4) WHERE IT WAS VISUALLY INSPECTED. THE DAMAGED COMPONENTS WERE THEN SENT TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR FURTHER INVESTIGATION. RESULTS: VISUAL INSPECTION REVEALED THAT THE GAS INLET PORT WAS CRACKED AND THE GAS OUTLET PORT WAS BROKEN ON THE RETURNED NEOPUFF DEVICE. THE UPPER END CAP WAS ALSO BROKEN. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT 030623. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL WARNS AGAINST DROPPING THE NEOPUFF OR SUBJECTING IT TO IMPACT DAMAGE WHICH MAY CAUSE THE UNIT TO OPERATE INCORRECTLY. IF THE NEOPUFF IS SUSPECTED TO HAVE BEEN DAMAGED, THE MANOMETER AND VALVE SYSTEM SHOULD BE PERFORMANCE TESTED. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT." THE COMPLAINT NEOPUFF DEVICE WAS FITTED WITH A NEW FASCIA, VALVE SYSTEM AND END CAP. THE NEOPUFF WAS RETURNED TO THE CUSTOMER AFTER PASSING THE SAFETY AND PERFORMANCE TESTS.
A HOSPITAL IN (B)(6) REPORTED THAT THE GAS INLET & OUTLET PORTS OF AN RD900 NEOPUFF INFANT RESUSCITATOR WERE BROKEN. A SERVICE OF THE DEVICE WAS REQUESTED.NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED THAT THE GAS INLET AND OUTLET PORTS OF AN RD900 NEOPUFF INFANT RESUSCITATOR WERE BROKEN. A SERVICE OF THE DEVICE WAS REQUESTED. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401203 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 030623 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |