FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3920831 · Received July 8, 2014

Report

Report Number
3004209178-2014-86565
Event Type
Injury
Date Received
July 8, 2014
Date of Event
June 10, 2014
Report Date
June 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UNIT PASSED ALL FUNCTIONAL TESTING INCLUDING THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. NO EXCESSIVE NO DELIVERY ALARMS NOTED. NO UNEXPECTED BATTERY OUT LIMIT ALARMS WERE NOTED. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS AND A SCRATCHED LCD WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE HAS BEEN HOSPITALIZED THREE TIMES DUE TO DIABETES KETOACIDOSIS AND HAS DISCONTINUED THE USE OF THE INSULIN PUMP SINCE THE EVENTS. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE HOSPITALIZATION WAS OVER 500 MG/DL IN ALL THREE INSTANCES. TROUBLESHOOTING WAS PERFORMED AND THE DRIVE SUPPORT CAP AND ALL SETTINGS APPEARED TO BE NORMAL. THE HIGH PRESSURE TEST WAS ALSO PERFORMED AND PASSED. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 104 MG/DL AND HAS TREATED WITH A MANUAL INJECTION. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398331 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization