PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3920823 · Received July 8, 2014

Report

Report Number
3004209178-2014-86540
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP UNABLE TO PRIME DURING PRIME TEST. AN A33 ALARMED DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK. INSULIN PUMP RECEIVED WITH MINOR SCRATCH ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS CRACKED, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED RECEIVING AN ALARM ON THE INSULIN PUMP AND INSULIN FLOWING OUT. THE BLOOD GLUCOSE WAS 80 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE DEVICE AND CHECK THE DRIVE SUPPORT CAP. UPON INSPECTION THE CUSTOMER NOTICED THAT IT WAS PROTRUDED. THE CUSTOMER WAS ADVISED TO DISCONTINUE THE USE OF THE DEVICE AND TO REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398446 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 55 YR