FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3920803 · Received July 8, 2014

Report

Report Number
2955842-2014-04169
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
May 29, 2014
Report Date
June 11, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE TIPS OF THE INSTRUMENTS GRIPS WERE BENT. ONE GRIP WAS BENT, CAUSING A SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .030 OFFSET AT THE TIP. THE BENT GRIP HAD SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT INDICATING OVERLOADING. FAILURE ANALYSIS CONCLUDED THE DAMAGE MAY BE DUE TO MISHANDLING / MISUSE. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. AN ADDITIONAL FINDING WAS VARIOUS SCRATCHES ON THE DISTAL END OF THE MAIN TUBE EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT AXIALLY ALIGNED WITH THE TUBE AXIS. AN ADDITIONAL FINDING WAS INSULATION ON THE CONDUCTOR WIRE WAS DAMAGED AT THE YAW PULLEY. THERE WAS NO EVIDENCE OF ARCING VISIBLE. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND PASSED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI PROSTATECTOMY PROCEDURE, THE JAWS WERE NOT ALIGNED ON THE MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398312 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 N10140109 098

Patients

Seq Age Sex Outcome Treatment
1