MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04169
- Event Type
- Malfunction
- Date Received
- July 8, 2014
- Date of Event
- May 29, 2014
- Report Date
- June 11, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE TIPS OF THE INSTRUMENTS GRIPS WERE BENT. ONE GRIP WAS BENT, CAUSING A SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .030 OFFSET AT THE TIP. THE BENT GRIP HAD SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT INDICATING OVERLOADING. FAILURE ANALYSIS CONCLUDED THE DAMAGE MAY BE DUE TO MISHANDLING / MISUSE. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. AN ADDITIONAL FINDING WAS VARIOUS SCRATCHES ON THE DISTAL END OF THE MAIN TUBE EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT AXIALLY ALIGNED WITH THE TUBE AXIS. AN ADDITIONAL FINDING WAS INSULATION ON THE CONDUCTOR WIRE WAS DAMAGED AT THE YAW PULLEY. THERE WAS NO EVIDENCE OF ARCING VISIBLE. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND PASSED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI PROSTATECTOMY PROCEDURE, THE JAWS WERE NOT ALIGNED ON THE MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 398312 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | N10140109 098 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |