FDA Adverse Event Malfunction Summary report: N

ACCENT DR RF

MDR report key: 3920780 · Received July 8, 2014

Report

Report Number
2017865-2014-14385
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
April 15, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED TO THE CLINIC FOR FOLLOW UP, THE PULSE GENERATOR EXHIBITED BACKUP VVI MODE. AFTER A DEVICE SOFTWARE DOWNLOAD, NORMAL DEVICE FUNCTION RESUMED. THE PATIENT WOULD BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396094 ACCENT DR RF IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2210 NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR