FDA Adverse Event Malfunction Summary report: N

SUSTAIN XL DR

MDR report key: 3920698 · Received July 8, 2014

Report

Report Number
2017865-2014-14311
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
May 9, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LEAD REPOSITIONING PROCEDURE, THE PULSE GENERATOR EXHIBITED BACKUP VVI MODE AFTER SUSPECTED ELECTROCAUTERY INTERACTION. AFTER A DEVICE SOFTWARE DOWNLOAD, THE DEVICE RESUMED NORMAL FUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396899 SUSTAIN XL DR IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2136 NA

Patients

Seq Age Sex Outcome Treatment
1