ENDURITY DR

MDR report key: 3920628 · Received July 8, 2014

Report

Report Number
2017865-2014-14232
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
March 28, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
P880086
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS CONFIRMED THAT IT WAS DIFFICULT TO INSERT THE TEST LEAD(S) INTO THE PULSE GENERATORS HEADER AND THE LEAD INSERTION FORCE USED TO INSERT THE LEAD WAS OUT OF SPECIFICATION FOR THE PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT IT WAS DIFFICULT TO INSERT THE ATRIAL LEAD INTO THE PULSE GENERATOR HEADER; IT DID NOT GO ALL THE WAY IN. THE DEVICE WAS NOT USED AND A NEW DEVICE WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398189 ENDURITY DR IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2160 NA