FDA Adverse Event Injury Summary report: N

ENDURITY DR

MDR report key: 3920616 · Received July 8, 2014

Report

Report Number
2017865-2014-14214
Event Type
Injury
Date Received
July 8, 2014
Date of Event
May 2, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
P880086
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND THE DEVICE IN BACKUP VVI MODE DUE TO HIGH CURRENT DRAIN, CAUSED BY A LEAKY C4 CAPACITOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED BACKUP OPERATION POST IMPLANT. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396103 ENDURITY DR IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2160 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention