ENDOTAK RELIANCE
MDR report key: 3920365
·
Received July 8, 2014
Report
- Report Number
- 2124215-2014-13648
- Event Type
- Injury
- Date Received
- July 8, 2014
- Date of Event
- June 12, 2014
- Report Date
- June 12, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE LEAD WAS NOT RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT HAD TWIDDLERS SYNDROME AND THIS LEAD WAS EXHIBITING IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS VISIBLY TWISTED AND THE SLACK IN THE LEAD WAS GONE. THE LEAD WAS REMOVED FROM SERVICE AND REPLACED WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 397776 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0293 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| L | E142| 4471| 0293 |