FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3920358 · Received July 8, 2014

Report

Report Number
2955842-2014-04167
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
June 11, 2014
Report Date
June 17, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THAT BOTH THE PITCH UP AND DOWN CABLES WERE FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE AND/OR TUBE ABRASIONS WITH MATERIAL REMOVED, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, A WIRE SEPARATED ON THE PROGRASP FORCEPS INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397512 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-11 N10140423 580

Patients

Seq Age Sex Outcome Treatment
1