FDA Adverse Event Malfunction Summary report: N

BATT-HANDPIECE MOD F/TRS

MDR report key: 3920149 · Received July 8, 2014

Report

Report Number
8030965-2014-00448
Event Type
Malfunction
Date Received
July 8, 2014
Report Date
September 27, 2012
Manufacturer
SYNTHES GMBH
Product Code
GEY
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE OR UNCHANGED. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. (B)(6). DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, NO FAILURE WAS IDENTIFIED IN THE FUNCTIONAL TESTING. THE UNIT WAS SERVICED AND RETURNED TO THE CUSTOMER ON (B)(6) 2012. PLACEHOLDER.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE BATTERY HANDPIECE, MODULAR, FOR TRAUMA RECON SYSTEM OVERHEATS. THIS IS REPORT 1 OF 1 FOR EVENT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397140 BATT-HANDPIECE MOD F/TRS GEY SYNTHES GMBH 5668

Patients

Seq Age Sex Outcome Treatment
1