FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3920131 · Received July 8, 2014

Report

Report Number
1416980-2014-21856
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
June 13, 2014
Report Date
June 13, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THE DEVICE PASSED ELECTRICAL AND FUNCTIONAL TESTING. EXTERNAL AND INTERNAL VISUAL INSPECTIONS WERE PERFORMED AND FOUND NO ISSUES. THE PNEUMATIC SYSTEM WAS TESTED AND FOUND NO ISSUES. A SIMULATED THERAPY WAS PERFORMED ON THE DEVICE SUCCESSFULLY. THE REPORTED ISSUE WAS IDENTIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS A FALSE EMPTY DETECTED AND AN INAPPROPRIATELY SET MINIMUM DRAIN VOLUME PERCENT SETTING TOO LOW. THE HOMECHOICE APD SYSTEMS TRAINER'S GUIDE PROVIDES A WARNING "IF YOU SET THE MINIMUM DRAIN VOLUME % TOO LOW, AN INCOMPLETE DRAIN COULD RESULT FOR THE PATIENT. THIS COULD RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION." A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HIGH DRAIN 105 ALARM OCCURRED DURING DRAIN FIVE ON THE HOME CHOICE (HC). THIS INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ARRANGED THE RETURN OF THE DEVICE AND DISCUSSED THE USE OF MANUAL SUPPLIES TO COMPLETE THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397121 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 83 YR