FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3920031 · Received July 8, 2014

Report

Report Number
3004209178-2014-86490
Event Type
Malfunction
Date Received
July 8, 2014
Date of Event
June 16, 2014
Report Date
June 16, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. UNABLE TO VERIFY THE ERROR ALARMS DUE TO MOTOR ERROR ALARMS. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS, AND MINOR SCRATCHED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THE INSULIN PUMP HAD BEEN ALARMING WHILE TRYING TO PLACE RESERVOIR INTO THE DEVICE. DURING TROUBLESHOOTING IT WAS FOUND THAT THE DRIVE SUPPORT CAP WAS PROTRUDING FROM THE DEVICE. CUSTOMER STATES THEY HAD NOT DROPPED THE DEVICE OR TRIED TO PUSH THE DRIVE SUPPORT CAP INTO A RECESSED POSITION. ADVISED CUSTOMER TO DISCONNECT FROM THE INSULIN PUMP AND INFORMED THAT THE DEVICE ALARMING WAS DUE TO THE FORCE SENSOR BEING UNABLE TO DETECT THE RESERVOIR. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO HEALTHCARE PROVIDER'S INSTRUCTIONS. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 98 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396492 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 32 YR