FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3920011 · Received July 8, 2014

Report

Report Number
2531779-2014-19348
Event Type
Malfunction
Date Received
July 8, 2014
Report Date
June 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP KEYPAD WAS OBSERVED TO BE PEELING AT THE OK BUTTON; THE KEYPAD RESPONSE WAS TESTED AND FOUND THAT THE CONTRAST BUTTON WAS INTERMITTENTLY RESPONDING BUT ALL OTHER BUTTONS RESPONDED APPROPRIATELY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS. PRODUCT ANALYSIS EVALUATED THE PUMP AND FOUND THAT CONTAMINATION WAS PRESENT UNDER THE KEYPAD BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397033 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 17 YR