FDA Adverse Event Other Summary report: N

C-FLEX ASPHERIC

MDR report key: 3919546 · Received June 10, 2014

Report

Report Number
9611165-2014-00056
Event Type
Other
Date Received
June 10, 2014
Date of Event
May 20, 2014
Report Date
June 2, 2014
Manufacturer
RAYNER INTRAOCULAR LENSES LTD.
Product Code
HQL
PMA / PMN Number
P060011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

RAYNER INTRAOCULAR LENSES LIMITED HAS ALLOCATED THE REF: (B)(4) TO THIS CASE. THE C-FLEX ASPHERIC 970C WAS LOADED BY THE SCRUB NURSE. THE HEALTHCARE FACILITY REPORTS THAT NO RESISTANCE WAS ENCOUNTERED DURING INJECTION. HAPTIC BREAKAGE WAS OBSERVED UPON IMPLANTATION OF THE IOL INTO THE EYE. REMEDIAL, CORRECTIVE AND PREVENTIVE ACTION WAS TAKEN BY THE HEALTHCARE FACILITY IN THE ORIGINAL PLANNED SURGERY SESSION. THE C-FLEX ASPHERIC 970C IOL SUBJECT TO THIS REPORT WAS EXPLANTED. A BACK-UP IOL WAS IMPLANTED WITHOUT FURTHER COMPLICATION AND WITHOUT ANY ADVERSE EFFECT TO THE PT. THE INFO RECEIVED INDICATES THAT HAPTIC BREAKAGE MAY HAVE OCCURRED AS A RESULT OF A USER LOADING ERROR. A MEDICAL ACCUJECT INJECTOR WAS UTILISED TO IMPLANT THE C-FLEX ASPHERIC 970C IOL. THE INJECTOR WAS RETAINED AND RETURNED TO RAYNER. RAYNER HAS SENT THE ACCUJECT INJECTOR TO THE MANUFACTURER (MEDICEL AG) FOR INSPECTION. THE C-FLEX ASPHERIC 970C IOL WAD DISCARDED BY THE HEALTHCARE FACILITY FOLLOWING EXPLANTATION. NO DEVICE ANALYSIS CAN BE UNDERTAKEN BY RAYNER IN THIS INSTANCE. OUR REVIEW OF PRODUCTION RECORDS FOR THE C-FLEX ASPHERIC 970C IOL BATCH 123E54282 SHOWED THAT ALL MANUFACTURING AND QUALITY CHECKS WERE CONDUCTED WITH SUCCESSFUL RESULTS. ALL LENSES RELEASED FOR DISTRIBUTION FROM THIS BATCH WERE WITHIN TOLERANCE, MET SPECIFICATION CRITERIA AND WERE WITHOUT DEFECTS.

Description of Event or Problem · 1

RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A (B)(6) HEALTHCARE FACILITY OF AN EVENT THAT OCCURRED DURING IMPLANTATION OF A C-FLEX ASPHERIC 970C IOL. THE EVENT DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY STATES "LEADING HAPTIC WAS BROKEN WHEN LENS WAS PLACED INTO THE EYE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338255 C-FLEX ASPHERIC HQL: INTRAOCULAR LENS HQL RAYNER INTRAOCULAR LENSES LTD. 970C 123E54282