FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3915898 · Received June 12, 2014

Report

Report Number
1720753-2014-04974
Event Type
Malfunction
Date Received
June 12, 2014
Date of Event
May 26, 2014
Report Date
June 12, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE FSE REPORTED THAT THE SYSTEM LOCKED UP AND CONTINUED TO BEEP AFTER THE EXPOSURE BUTTON WAS RELEASED. THIS COULD RESULT IN THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
347257 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1