FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3915850
·
Received June 12, 2014
Report
- Report Number
- 1720753-2014-05011
- Event Type
- Malfunction
- Date Received
- June 12, 2014
- Report Date
- June 12, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE RIBBON CABLE WAS RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE MONITORS HAD BLACK AND WHITE VERTICAL STRIPS, RESULTING IN THE DISPLAYED IMAGES BEING UNUSABLE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347521 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|