FDA Adverse Event
Injury
Summary report: N
ZEPHYR XL DR
MDR report key: 3915776
·
Received January 15, 2014
Report
- Report Number
- 2017865-2014-08556
- Event Type
- Injury
- Date Received
- January 15, 2014
- Date of Event
- March 18, 2013
- Manufacturer
- ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWP
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFO PROVIDED BY MFR, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVAL DESCRIPTION INCLUDED. ANALYSIS CONFIRMED NORMAL BATTERY DEPLETION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED PREMATURE BATTERY DEPLETION. DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 39026 | ZEPHYR XL DR | IMPLANTABLE PACEMAKER PULSE GENERATOR | LWP | ST. JUDE MEDICAL, CARDIAC RHYTHM MANAGEMENT DIVISION | 5820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention |