HOMECHOICE PRO
MDR report key: 3915749
·
Received July 7, 2014
Report
- Report Number
- 1416980-2014-21653
- Event Type
- Malfunction
- Date Received
- July 7, 2014
- Date of Event
- April 24, 2014
- Report Date
- June 12, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE OF THE IIPV EVENT COULD NOT BE DETERMINED WITH THE AVAILABLE INFORMATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
A HIGH DRAIN ERROR 106 (NIGHT DRAIN #4) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE (HC) DEVICE. A HIGH DRAIN ALARM OCCURS WHEN A PATIENT HAS DRAINED A VOLUME EQUAL TO 200% OR GREATER THAN THE PATIENTS FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2014 AT 06:07:13. NO ADDITIONAL INFORMATION IS AVAILABLE.