FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3915749 · Received July 7, 2014

Report

Report Number
1416980-2014-21653
Event Type
Malfunction
Date Received
July 7, 2014
Date of Event
April 24, 2014
Report Date
June 12, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. THE CAUSE OF THE IIPV EVENT COULD NOT BE DETERMINED WITH THE AVAILABLE INFORMATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

A HIGH DRAIN ERROR 106 (NIGHT DRAIN #4) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE (HC) DEVICE. A HIGH DRAIN ALARM OCCURS WHEN A PATIENT HAS DRAINED A VOLUME EQUAL TO 200% OR GREATER THAN THE PATIENTS FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2014 AT 06:07:13. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
394009 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1