ETS FLEX ARTICNG LNR CUTR 45MM
Report
- Report Number
- 3005075853-2014-04677
- Event Type
- Malfunction
- Date Received
- July 7, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 16, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS: DAMAGED SPRING CARTRIDGE LOCKOUT TAB. THE ANALYSIS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A CARTRIDGE LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED 1/10 AND WITH THE RELOAD LOCK OUT SPRING DAMAGED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE NURSE IN THE ROOM SAID "IT WOULD NOT FIRE." IT WAS THE FIRST FIRING. THERE WERE NO PATIENT CONSEQUENCES. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 393178 | ETS FLEX ARTICNG LNR CUTR 45MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |