ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19201
- Event Type
- Malfunction
- Date Received
- July 7, 2014
- Report Date
- June 29, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/19/2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH NO VISIBLE DAMAGE TO THE KEYPAD COVER. DURING TESTING, ALL OF THE KEYPAD BUTTONS WERE NORMALLY RESPONSIVE. THE KEYPAD COVER WAS REMOVED AND NO CONTAMINATION WAS FOUND UNDER ANY OF THE KEYPAD BUTTON CONTACTS. THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. UNRELATED TO THE INITIAL COMPLAINT, THERE WAS EVIDENCE OF MOISTURE FOUND INSIDE THE PUMP. ALSO UNRELATED, THERE WERE NO AUDIBLE TONES DUE TO THE CRACKED SOLDER ON PIEZO. THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES W/MOISTURE) ISSUE. REPORTEDLY, THE ¿UP¿, ¿DOWN¿ AND ¿OK¿ BUTTONS WERE UNDER RESPONSIVE AND MOISTURE WAS OBSERVED IN THE BATTERY COMPARTMENT. IT WAS REPORTED THAT THERE WAS NO DAMAGE TO THE KEYPAD COVER, AND THERE WAS NO DELIVERY ISSUES NOTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 393598 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |