FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3915338 · Received July 7, 2014

Report

Report Number
2531779-2014-19189
Event Type
Malfunction
Date Received
July 7, 2014
Report Date
June 23, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 10/06/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING A VISUAL INSPECTION OF THE PUMP, NO DAMAGE WAS FOUND TO THE KEYPAD COVER. DURING TESTING, ALL OF THE KEYPAD BUTTONS RESPONDED PROPERLY. THE KEYPAD COVER WAS REMOVED, AND CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS. THE DOWN ARROW BUTTON CONTACT WAS ALSO DAMAGED AND DIDN¿T SPRING BACK AFTER BEING PRESSED. ADDITIONALLY, THE BOLUS BUTTON COVER WAS TORN; THE BUTTON RESPONDED PROPERLY. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED. ADDITIONALLY, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT WITH THIS COMPLAINT. THERE WAS NO FURTHER INFORMATION AVAILABLE REGARDING THE COMPLAINT AT THIS TIME; IF FURTHER INFORMATION IS PROVIDED A FOLLOW UP REPORT SHALL BE MADE. THIS COMPLAINT IS BEING REPORTED BECAUSE OF THE ALLEGATION OF A KEYPAD BUTTON RESPONSE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
394231 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 42 YR