840 VENTILATOR
MDR report key: 3915300
·
Received May 27, 2014
Report
- Report Number
- 8020893-2014-01275
- Event Type
- Malfunction
- Date Received
- May 27, 2014
- Date of Event
- January 1, 2014
- Report Date
- April 29, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARD (PCBS). THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).
Description of Event or Problem · 1
COVIDIEN RECEIVED INFORMATION FROM A CUSTOMER IN THE USA STATING THAT AN 840 VENTILATOR FAILED POWER ON SELF TEST (POST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE EXACT DATE OF THE EVENT IS UNKNOWN.