FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3915300 · Received May 27, 2014

Report

Report Number
8020893-2014-01275
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
January 1, 2014
Report Date
April 29, 2014
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARD (PCBS). THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION FROM A CUSTOMER IN THE USA STATING THAT AN 840 VENTILATOR FAILED POWER ON SELF TEST (POST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED. THE EXACT DATE OF THE EVENT IS UNKNOWN.